![]() | RAD-Check Plus X-Ray Test Instrument. | ![]() |
![]() |
Key FeaturesIntroduction Proven RAD-CHECK technology specifically designed to provide you with the ultimate in versatility and cost-effective operation. Accurate, lightweight, portable...this “industry standard” enables you to gain the critical edge in your QC program. Applications Battery operation and built-in detector virtually eliminate setup time. Just place the meter or external ion chamber on |
| x-ray table; collimate, shoot, and read the result. Precision ion chamber and digital display ensure accuracy plus easy
readability. Specifications Ranges 0.001 to 2 R, 0.01 to 20 R/min Internal chamber 30 cc volume, energy response ± 5% from 15 to 65 keV (30 to 150 kVp filtered). 20.5 cm² (5.1 cm Ø) effective measurement area. Center of chamber 1.03 cm below top of chamber Standard calibration At 75 kVp with 4 mm Al filtration at 22°C and one atmosphere Reproducibility Within 2% short-term over 100 mR to 2 R range (1 to 20 mGy) Electrometer drift 0.5 to 1 mR/min typical; 6 mR/min maximum (5 to 10 μGy; 60 μGy/min maximum) Maximum exposure rate Minimum 90% collection efficiency at 20 R/sec Automatic reset Resets display to zero; can also be reset manually Operating temperature 10 to 40°C Relative humidity 0 to 90%, non-condensing Display 3.50 x 0.50 inch LCD, low battery indicator Controls Auto or manual reset selector. Display zero reset button. Dose or dose rate output selector. Integral or remote ion chamber selector. On/off switch Power 9 V alkaline battery, > 100 hours operation (50 hours in manual reset mode) Dimensions 6 (w) x 6.25 (d) x 2.75 in (h) (15.25 x 15.9 x 7 cm) Weight 1.125 lb (0.51 kg) |
|
|
![]() |
![]() |