Oprax Medical Inc.

Precision High-Voltage Divider.


Fluke Biomedical, Inc.
  • Frequency response is within 2 dB DC, to 100 kHz
  • Voltage rating anode to cathode: 160 kVp
  • Includes frequency compensation circuit
  • Conservative design
  • Generous insulation

  • Introduction
    The Wave views the anode and cathode waveforms of x-ray tubes at levels up to 150 kVp. Its frequency response is above 100 kHz. The circuit requires no external power and has a ratio of 100,000:1, 100 kVp = 1.0 kVp.

    Applications
  • Checking the waveform flatness in cine or photofluoroscopy imaging
  • Observing the discharge waveform of capacitive mobile x-ray machines
  • Examining the voltage compensation of falling load machines
  • Checking mammography voltage smoothing and checking for defective contactors
  • Checking for spiking at leading edge of waveforms (cable killers)


  • Checking voltage commutation of energy switchers (bone cancelling DSA)

  • Specifications
    Voltage rating
    - Anode to cathode 160 kVp
    - Anode to ground 80 kVp
    - Cathode to ground 80 kVp
    Ratio 100,000: 1, 2% or better
    Accuracy ± 2% or better
    Frequency response Within 2 dB DC, to 100 kHz
    Output impedance 1,000 ohms, 1%
    Connector Federal standard
    Output Anode-Cathode BNC type
    Insulation Oil
    Dimensions 11 in Ø x 12 in (h) (28 x 30.5 cm)
    Weight 35 lb (16 kg)


    Stock #
    Description
    Your Price
    07-469
    The Wave Precision High-Voltage Divider, without cables
    $3,044.29
    07-469-4780
    The Wave Precision High Voltage Divider, with 2 cables
    $3,951.40
    07-478
    High-Voltage Cable, 5 ft.
    $642.20
    89-476
    Carrying Case
    $450.04