Oprax Medical Inc.

Primalert Digital Radiation Area Monitors.


Fluke Biomedical, Inc.
  • LED digital display
  • Low and High alarm indicators
  • Programmable alarm indicator
  • Optional remote
  • Detector Fail indicator
  • Battery backup
  • Data output/RS-232
  • Two configurations available with internal energy compensated GM detectors:
  • - 0.1 mR/hr to 1R/hr (Model 05-443)
    - 1 mR/hr to 4 R/hr (Model 05-444)
  • CE tested. Meets applicable standards

  • Features
  • Siimple installation and setup
  • Anti-jam circuitry prevents erronious readings at tube saturation
  • Calibration controls easily accessed through front panel

  • Introduction

    The PRIMALERT Digital Area Monitors are designed for a wide range of radiation area monitoring applications. They are completely self-contained units with energy compensated GM detectors, AC powered with internal battery backup and have user settable low and high alarms. An optional remote alarm is available for added security.

    Applications

    The PRIMALERT Digital Area Monitors can be used wherever there is a need to detect and warn personnel of increasing radiation levels and to control the exposure of personnel to gamma radiation. The versatility of the PRIMALERT allows it to be used from industrial applications to medical settings.


    Specifications

    Detector
    Indicated use Radiation area monitoring
    Internal GM detector range: Model 05-443 0.1 mR/hr to 1 R/hr; Model 05-444 1 mR/hr to 4 R/hr
    Display 4 digit LED display with .8 in (2 cm) character height
    Display range 000.0 to 9999
    Display units Can be made to display in µR/hr, mR/hr, R/h, µSv/h, mSv/h, Sv/h, cpm, cps and others
    Linearity Reading within ±10% of true value with detector connected
    Response Typically 3 seconds from 10% to 90% of final reading
    Status (green light) Indicates the instrument is functioning properly
    Low alarm Indicated by a yellow light and slow beep (1 per sec) audible tone (can be set at any point from 0.0 to 9999)

    Note; audible indicators can be configured as a single beep if desired

    Detector fail (red light and audible tone; > 68 dB at 2 ft) Indicates detector overload, no count from detector, or instrument failure
    Low battery (yellow) Indicates < 2 hours of battery power remaining
    Calibration controls Accessible from front of instrument (protective cover provided)
    High voltage Adjustable from 200 to 2500 V
    Threshold Adjustable from 2 to 100 mV
    Dead time Adjustable to compensate for dead time of the detector and electronics (can be read on the display)
    Overload Senses detector saturation (indicated by display reading "-OL")
    Overrange Indicates the radiation field being measured has exceeded the counting range of the instrument (indicated by display reading "----")
    Data output 9 pin connector providing 5 decade log output, RS-232 output, signal ground connection FAIL and Alarm signals (current sink), and direct connection to battery and ground


    Power requirements 95 to 135 VAC (178 to 240 VAC available), 50 to 60 Hz single phase (< 100 mA), 6V sealed lead acid rechargeable battery (built-in)
    Battery Life Typically 48 hours in non-alarm condition, 12 hours in alarm condition
    Battery charger Battery is continuously trickle charged when instrument is connected to line power and turned on
    Construction Aluminum housing with white polyurethane enamel paint
    Temperature raange -4º to 122ºF (-20º to 50ºC). May be certified for operation from -40º to 150ºF (-40º to 65ºC)
    Size 9.7"W x 2.5"D x 7.4"H (24.6 x 6.4 x 18.7 cm)
    Weight 6.5 lb (2.3 kg)

    Optional remote display (Model 05-446)
    Indicated use Remote display for PRIMALERT Area Monitors Models 05-443 and 05-444
    Status (green light) Indicates the instrument is functiong properly
    High alarm (red LED) Indicates the radiation level exceeds the high alarm point
    Detector fail (red LED) Indicates detector overload, no count from detector, or instrument failure
    Audio Unimorph type with ON/OFF switch (< 68 dB at 2 ft)
    Power requirements provided by PRIMALERT (20 ft. cable included)
    Construction Aluminum housing with ivory polyurethane enemel point
    Temperature range -4º to 122ºF (-20º to 50ºC). May be certified for operation from -40º to 150ºF (-40º to 65ºC)
    Size 6"W x 2"D x 7"H (15.2 x 5.1 x 17.8 cm)
    Weight 1.5 lb (0.7 kg)

    Available models
    05-443 PRIMALERT Digital Area Monitor with internal energy compensated 0.1 mR/hr to 1 R/hr GM detector
    05-444 PRIMALERT Digital Area Monitor with internal energy compensated 1 mR/hr to 4R/hr GM detector

    Optional accessories
    Remote Display (Model 04-446)



    Stock #
    Description
    Your Price
    05-443
    PRIMALERT Digital Area Monitor w/Internal energy compensated 0.1 to 1000 mR/hr GM detector
    $2,174.63
    05-444
    PRIMALERT Digital Area Monitor w/internal energy compensated 1 to 4000 mR/hr GM detector
    $2,056.31
    05-446
    PRIMALERT Digital Area Monitor Remote Alarm
    $898.53