Specifications
| Description | Features | Benefits |
| Applications | Ergonomically designed for superior operation during portable X-ray applications | Used in hospitals, medical centers, clinics, sports complexes, nursing homes, home healthcare, correctional facilities & military applications. |
| Generator Type | High Frequency Inverter Microprocessor Controlled | 1.25 kW Regulated for both mA & kV |
| kVp Range | 50-100 | Adjustable in 1 kVp steps for optimum resolution |
| mA | 12.5 | Constant independent of kVp or time settings |
| mAs Range | 0.125-50 | Optimum mAs range |
| Exposure Time | 0.01-4.0 seconds | 96 steps for accurate technique programming |
| X-Ray Tube | 15 degree Target Stationary Anode 25,000 Heat Unit Storage | Full 360 degree horizontal rotation, 270 degree tube axis rotation |
| Focal Spot Target | 1.0 mm (NEMA) | Maximum resolution for superior image quality |
| Collimator | D-180HS-G | 360 degree rotation with high output halogen lamp |
| Exposure Switch | 2 Position (Prep-Exp) | 12 ft. long cable |
| Cassette Storage | Included | Stores up to 7 @ 14 x 17's |
| Tube/Column Locks | Heavy Duty/Manual | Friction type for fast & easy operation |
| Braking System | Heavy Duty/Lever Type Manual | Assures proper position is maintained |
| System Dimensions |   | |
| Operational Mode:25"W x 43"L x 52"H | Easily maneuvered, stored & transported | |
| Storage Mode:23"W x 53"L x 27"H | 2-step folding procedure | |
| SID | 15" - 78" from floor | Optimum SID range |
| Wheels | 12" inflatable | Soft for easy transportation |
| Weight | 95 lbs. | Lightweight-User friendly |
| Electrical Specifications | 103-126 VAC, 50/60 HZ | Broad operational range |
| 207-253 VAC, 50/60 HZ | Plugs into regular wall outlet |
| Product Compliance | FDA 510(k), UL, CUL, CE, TUV, IEC601-1 & 1-2-7 | Certified to 21 CFR subchapter 21 Electromagnetic compatible |
| | |