![]() | Near Field Ultrasound Phantom. | ![]() |
| (1) US PATENT# 5196343 Material:
Zerdine® 1 solid elastic water-based polymer. Freezing Point: 0°C Melting Point: Above 100°C Attenuation Coefficient: 0.50 dB/cm-MHz Velocity of Sound Through Background Medium: 1540 m/s ± 10 m/s Scanning Well: 6.5" X 3.9" X 0.6" (165mm X 100mm X 15mm) Scanning Membrane: Saran-based laminate. Targets:
Material: Stainless Steel Wire Diameter: 0.1mm Vertical Plane Targets:
Depth Range: 8cm Spacing: 1cm | ![]() |
| Axial Resolution Targets:
Number of Arrays: 2 Depths: 1.5cm and 2cm Axial Intervals: 0.5, 1, 2, 3, 4, and 5mm Horizontal Intervals: 1, 2, 3, 4, and 5mm Ring Down Target:
1mm to 10mm Volumetric Test Object:
Calibrated asymmetric shape 7cc and 21cc Spherical Cysts:
Diameter: 5mm, 3mm Random Distribution High Contrast/Cystic:
Cylinder Diameter: 10mm Depth: 1.0cm Contrast: Hyperechoic/Cyst-like |
![]() |
![]() |