![]() | General Purpose Urethane Ultrasound Phantom. | ![]() |
| Visualized Spacing: 20.0mm Material: 0.1mm ø, Nylon mono-filament. Horizontal Plane Targets:
Number of Groups: 1 Number of Targets Per Group: 10 Depth of Visualization: 3 & 10 cm Visualized Spacing: 18.6mm Material: 0.1mm ø, Nylon mono-filament. Note: This target group is also the Vertical Plane Target group. Axial Resolution Targets:
Number of Groups: 2 Number of Targets Per Group: 12 Depth of Visualization: 2, 5, 8 & 11 cm. Axial Resolution Test Range: 0.50mm, 1.0mm - 5.0mm in 1.0mm increments. Material: 0.1mm ø, Nylon mono-filament. Lateral Resolution Targets:
Number of Groups: 2 Number of Targets Per Group: 6 Depth of Visualization: 2, 5, 8 & 11 cm. Lateral Resolution Test Range: 1.0mm - 5.0mm in 1.0mm increments. Material: 0.1mm ø, Nylon mono-filament. Stepped Masses:
Number of Targets: 2 Diameter: 8mm - 2mm in 2mm increments. Depth of Visualization: 2, 5, 8, 11, 13 & 16 cm. Appearance: Cyst-like | ![]() |
![]() | General Purpose Ultrasound Phantom. | ![]() |
![]() |
The CIRS Model 054GS General Purpose Ultrasound Phantom offers all the features of the original Model 054 but now
contains gray scale targets and improved sensitivity targets that exceed the phantom requirements outlined in the ACR
accreditation program.The Model 054GS is constructed from the patented solid elastic material, Zerdine® (1). Zerdine® is a durable, reliable and accurate tissue mimicking material. The Model 054GS background material mimics the acoustic properties of soft tissue. A removable water well and endocavity cover extends the use of the phantom by allowing evaluation of all transducer configurations: linear, curvilinear and intercavity. CIRS ultrasound QA phantoms all come standard with a robust housing, rugged carry case, 4-year warranty, user guide and data sheet. (1) US PATENT# 5196343 Performance Measurements |
| Material:
Zerdine® 1 solid elastic water-based polymer. Freezing Point: 0°C Melting Point: Above 100°C Attenuation Coefficient: 0.50 dB/cm-MHz (0.7dB/cm-MHz available upon request) Velocity of Sound Through Background Medium: 1540 m/s ± 10 m/s Scanning Well: 5.9" X 3.9" X 0.4" (150mm X 100mm X 10mm) Scanning Membrane: Saran-based laminate. Targets:
Vertical Plane Targets:
Number of Groups: 1
Horizontal Plane Targets:Number Targets per Group: 7 Depth Range: 18cm Spacing: 2cm Material: 0.1mm Nylon monofilament
Number of Groups: 1
Number Targets per Group: 7 Depth Range: 9cm Spacing: 2cm Material: 0.1mm Nylon monofilament | ![]() |
| Near Field Targets:
Number of Groups: 1
Axial-Lateral Resolution Targets:Number Targets per Group: 6 Depth of Targets: 1, 2, 3, 4, 5, & 6mm Material: 0.1mm Nylon wire
Number of Groups: 2
Anechoic Cylinders:Number Targets per Group: 12 Depth of Targets: 3 & 11cm Spacing: 0.5mm, 1, 2, 3, 4, & 5mm Material: 0.8mm Nylon wire
Number of Targets: 5
Gray Scale Targets:Diameter of Target: 8mm Contrast: Anechoic
Number of Targets: 6
Diameter of Target: 8mm Depth of Target: 4cm Contrast: Anechoic, -6dB, -3dB, +3dB, +6dB hyperechoic. |
![]() | General Purpose, Multi-Tissue Ultrasound Phantom. | ![]() |
![]() |
The new standard for ultrasound quality assuranceThe CIRS Model 040GSE Multi-Purpose Multi-Tissue Ultrasound Phantom not only contains all the features of the original CIRS Model 040, but also includes gray scale targets, anechoic stepped masses and elasticity targets. The phantom is designed to meet the ultrasound QA challenges of today and tomorrow. The unique dual attenuation of the background gel allows for evaluation of transducers that range from 2 MHz - 15 MHz. A removable water well and endocavity cover extends the use of the phantom by allowing evaluation of all transducer configurations: linear, curvilinear and intercavity. CIRS ultrasound QA phantoms all come standard with a robust housing, rugged carry case, 4-year warranty, user guide and data sheet. Performance Measurements (1) US PATENT# 5196343 |
|
SPECIFICATIONS Material: Zerdine®(1), solid elastic water-based polymer. Freezing point: 00 C / Melting point: Above 1000 C Attenuation Coefficient: 0.5 dB/cm-MHz / 0.7 dB/cm-MHz Speed of Sound: 1540 m/s ± 10 m/s Scanning Well: 1 cm deep Scanning Membrane: Saran-based laminate NEAR FIELD GROUP:
Material: Nylon monofilament
HORIZONTAL AND VERTICAL GROUPS:Number of targets: 5 Depth range: 1 to 5 mm Distance between Targets: 1 mm
Material: Nylon monofilament
VERTICAL GROUP:
Wire diameter: 100 microns
Number of targets: 15
HORIZONTAL GROUPWire diameter: 100 microns Depth range: 1 to 16 cm Spacing: 10 mm
Number of groups: 2
Wire diameter: 100 microns Depth range: 4 and 9 cm Number of targets: 4 & 7 respectively Spacing: 10 & 20 mm respectively | ![]() |
![]() |
![]() |