MRI Multipurpose Phantom.


Fluke Biomedical, Inc. Introduction
The MRI Multipurpose Phantom monitors the overall performance of an MRI system. The parameters that can be measured include: slice thickness, slice orientation, interslice gap, magnetic field homogeneity, radio frequency signal uniformity, spatial resolution in positive and negative contrast, and modulation transfer function. The phantom can also be used to evaluate quadrature setting, baseline correction, aspect ratio, and single T1 and T2 values.

The phantom is constructed of non-magnetic materials. It is 9 inches in diameter, with two inserts, and is designed to be filled with an MRI signal-producing solution. One insert is a series of concentric conic sections. The other insert is made up of five sections: one for positive contrast, two for slice evaluation, one for MTF evaluation, one for T1 and T2 evaluations.

A void between the sections provides a fully flooded area for signal uniformity.



Key Features
The MRI Multipurpose Phantom provides a comprehensive range of tests in one compact unit. Tests include:

  • Slice thickness
  • MTF evaluation
  • Spatial resolution
  • RF signal uniformity
  • Magnetic field homogeneity
  • And many others

  • Product Specifications
    Standard Configuration by Section:
    Folded step ramps: two 60º
    Star pattern: one 120º
    T1, T2 solution insert: one 60º
    Flood section: one 360º
    Concentric conic section: one 360º
    Phantom dimensions: OD is 9 x 4.5 in (t)
    Weight: 6.82 lb (3.09 kg)




    Stock #
    Description
    Your Price
    76-903
    MRI Multipurpose Phantom
    $3,608.36
    76-903-7000
    Copper Sulfate, 1 gm/l
    $92.32